Sperm Counting ChamberDRM-600

Description

For sperm counts in undiluted specimens. The CELL-VU® Cytometer consists of a standard (3 x 1″), specially-designed glass slide and a coverslip with the counting grid laser-etched into its surface. Disposal of the entire apparatus eliminates cleaning and minimizes clinician exposure to body fluids. Both the slide and coverslip are marked to ensure correct use. The coverslip’s 1 x 1 mm grid is divided into 100, 0.1 mm areas and is thin enough to use under high magnification without special adapters or reticles. The slide consists of two chambers (for dual determinations), each with a depth of 20 µm. This depth is optimal for sperm cells to form in a monolayer, movement is unencumbered, motility can be assessed and counts are made easily.

25 Slides/50 Grid Coverslips

  Available for sales in the European Union

Specifications
Ordering Information
DRM600 | Disposable Sperm Counting Chambers

DRM600 - CELL-VU Sperm Counting Chamber Instructions for Use

 STEP 1

Ensure the CELL-VU® name on the cover glass is facing the observer as the grid is etched on the reverse side.

NOTE: Do NOT apply any chemicals or solvents to the slide or coverslip as doing so will affect the coating on the slide, and compromise the counting of sperm within the defined thickness.

STEP 2

Mix the sample thoroughly just before pipetting.

Pipette one drop (approximately 4 microliters) of specimen. Place the specimen at the extreme edge of one of the sampling areas (figure A).

STEP 3

Gently lower the cover glass over the specimen so that the edge of the cover glass just covers the sample (figure B).

STEP 4

Slide the cover glass into final position so that the circle on the cover glass is centered within the circle of the sampling area (Figures C and D).

This will eliminate air bubbles from the counting area.

STEP 5

SPERM CONCENTRATION

Isolate and view the CELL-VU® grid. This grid is divided into 100 small boxes each 0.1 x 0.1 mm.For undiluted sperm count all motile and non-motile sperm within 10 small boxes of this grid. Divide this number by 2. This result is the concentration of sperm in millions/ml. For specimens with low sperm counts, count all sperm within the entire grid (100 boxes). Multiply the count by 50,000 to obtain the total concentration of sperm/ml.

SPERM MOTILITY

Count both motile sperm and non-motile sperm within 10 small boxes of the grid. Divide total sperm (motile + non-motile) by the total motile count. Multiply by 100 to establish the percentage of motile sperm.